Search results for "3616 Speech and Hearing"
showing 1 items of 1 documents
Effect of changing pulse rate on profile parameters of perceptual thresholds and loudness comfort levels and relation to ECAP thresholds in recipient…
2010
Abstract: The Nucleus CI24RE Freedom device offers higher stimulation rates and lower noise levels in action potential measurements (ECAPs) than previous devices. A study including ten European implant teams showed that the effect of changes in rate from 250 to 3500 pulses per second on tilt and curvature of the T and C profiles is insignificant. When changing rate one may change the levels at all electrodes by the same amount. Using an automated procedure ECAPs could be measured quickly and reliably at a noise level of only 1 μV. However, this did not result in improved correlations between the tilt and curvature parameters of the ECAP profiles and those of the T and C profiles. Average C …